Skip to main navigation Skip to search Skip to main content

Determination of the dosimetric characteristics of InterSource125 Iodine brachytherapy source

Research output: Contribution to journal › Article › peer-review

29 Scopus citations

Abstract

The TG-43 recommended dosimetric characteristics of a new 125I brachytherapy source have been experimentally and theoretically determined. The measurements were performed in Solid Water™ using LiF TLDs. The calculations were performed using Monte Carlo simulations in Solid Water™ and water. The measured data were compared with calculated values as well as the reported data in literature for other source designs. The dose rate constant this source in water was 1.01±3% cGy h-1 U-1 and the anisotropy constant was 0.956.

Original languageEnglish
Pages (from-to)589-599
Number of pages11
JournalApplied Radiation and Isotopes
Volume56
Issue number4
DOIs
StatePublished - 2002

Funding

This project was partially supported by International Brachytherapy Inc. The authors would like to thank Mr. Mat M. Tamimi for his assitance during the TLD measurements and data analysis.

Funders
International Brachytherapy Inc.

    Keywords

    • I
    • Brachytherapy
    • Dosimetry
    • TG-43
    • Thermoluminescent dosimeter
    • TLD

    ASJC Scopus subject areas

    • Radiation

    Fingerprint

    Dive into the research topics of 'Determination of the dosimetric characteristics of InterSource125 Iodine brachytherapy source'. Together they form a unique fingerprint.