Abstract
The TG-43 recommended dosimetric characteristics of a new 125I brachytherapy source have been experimentally and theoretically determined. The measurements were performed in Solid Water™ using LiF TLDs. The calculations were performed using Monte Carlo simulations in Solid Water™ and water. The measured data were compared with calculated values as well as the reported data in literature for other source designs. The dose rate constant this source in water was 1.01±3% cGy h-1 U-1 and the anisotropy constant was 0.956.
| Original language | English |
|---|---|
| Pages (from-to) | 589-599 |
| Number of pages | 11 |
| Journal | Applied Radiation and Isotopes |
| Volume | 56 |
| Issue number | 4 |
| DOIs | |
| State | Published - 2002 |
Funding
This project was partially supported by International Brachytherapy Inc. The authors would like to thank Mr. Mat M. Tamimi for his assitance during the TLD measurements and data analysis.
| Funders |
|---|
| International Brachytherapy Inc. |
Keywords
- I
- Brachytherapy
- Dosimetry
- TG-43
- Thermoluminescent dosimeter
- TLD
ASJC Scopus subject areas
- Radiation
Fingerprint
Dive into the research topics of 'Determination of the dosimetric characteristics of InterSource125 Iodine brachytherapy source'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver