Dose verification with Monte Carlo technique for prostate brachytherapy implants with 125I sources

Hualin Zhang, Curtis Baker, Rachel McKinsey, Ali Meigooni

Research output: Contribution to journalArticlepeer-review

19 Scopus citations

Abstract

The Monte Carlo technique has been implemented to generate the dose distributions in a model prostate patient, implanted with iodine-125 ( 125I) brachytherapy sources. The results of these calculations are also compared with the dose distributions calculated by a commercially available treatment planning system. The comparison shows that with the source strength suggested by the Monte Carlo technique, the current clinical planning system is found to provide 100% coverage of the prostate with the prescription dose for the same implant pattern. However, the dose-volume histogram of this investigation shows that the VariSeed™ treatment planning system has a 29% and 136% larger dose coverage for the 150% and 200% isodose lines, respectively, than the Monte Carlo simulation. These differences are attributed to the oversimplification of the current planning system using the point source approximation, and also to the interseed effects from multisources that are neglected in the conventional planning systems. The results of this study provide evidence that supports the use of the Monte Carlo technique in treatment planning systems to provide accurate dose calculations in brachytherapy implants.

Original languageEnglish
Pages (from-to)85-91
Number of pages7
JournalMedical Dosimetry
Volume30
Issue number2
DOIs
StatePublished - 2005

Keywords

  • I source
  • Brachytherapy
  • Interseed effect
  • Monte Carlo
  • Prostate implant

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology
  • Oncology
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Dose verification with Monte Carlo technique for prostate brachytherapy implants with 125I sources'. Together they form a unique fingerprint.

Cite this