Skip to main navigation Skip to search Skip to main content

EchoSeed Model 6733 Iodine-125 brachytherapy source: Improved dosimetric characterization using the MCNP5 Monte Carlo code

  • M. A. Mosleh-Shirazi
  • , K. Hadad
  • , R. Faghihi
  • , M. Baradaran-Ghahfarokhi
  • , Z. Naghshnezhad
  • , A. S. Meigooni

Research output: Contribution to journalArticlepeer-review

15 Scopus citations

Abstract

This study primarily aimed to obtain the dosimetric characteristics of the Model 6733 125I seed (EchoSeed) with improved precision and accuracy using a more up-to-date Monte-Carlo code and data (MCNP5) compared to previously published results, including an uncertainty analysis. Its secondary aim was to compare the results obtained using the MCNP5, MCNP4c2, and PTRAN codes for simulation of this low-energy photon-emitting source. The EchoSeed geometry and chemical compositions together with a published 125I spectrum were used to perform dosimetric characterization of this source as per the updated AAPM TG-43 protocol. These simulations were performed in liquid water material in order to obtain the clinically applicable dosimetric parameters for this source model. Dose rate constants in liquid water, derived from MCNP4c2 and MCNP5 simulations, were found to be 0.993 cGyh-1 U-1 (±1.73) and 0.965 cGyh-1 U-1 (±1.68), respectively. Overall, the MCNP5 derived radial dose and 2D anisotropy functions results were generally closer to the measured data (within ±4) than MCNP4c and the published data for PTRAN code (Version 7.43), while the opposite was seen for dose rate constant. The generally improved MCNP5 Monte Carlo simulation may be attributed to a more recent and accurate cross-section library. However, some of the data points in the results obtained from the above-mentioned Monte Carlo codes showed no statistically significant differences. Derived dosimetric characteristics in liquid water are provided for clinical applications of this source model.

Original languageEnglish
Pages (from-to)4653-4659
Number of pages7
JournalMedical Physics
Volume39
Issue number8
DOIs
StatePublished - Aug 2012

Keywords

  • AAPM TG-43U1
  • brachytherapy
  • EchoSeed Model 6733 Iodine-125 source
  • MCNP
  • Monte Carlo simulation

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Fingerprint