Abstract

The purpose of this research is to study the effect of various applicator compositions on dosimetric parameters and dose distribution of 192Ir, 137Cs, and 60Co sources, using Monte Carlo simulation techniques. To study the effect of applicators on source dosimetry, the dose rate constant, and radial dose function and isodose curves for the above noted sources were calculated in the presence and absence of plastic, titanium, and a stainless steel applicators. The effects of the applicators on the dosimetric parameters and isodose curves of these sources were dependent of the source type and materials of the applicator. The 192Ir source with the stainless steel applicator has the maximum difference of dose rate (4.2 %) relative to the without applicator case. The 60Co source with plastic applicator has the minimum dose variation. Moreover, this effect is higher for lower energy sources. Ignoring the effect of applicator composition and geometry on dose distribution may cause discrepancies in treatment planning. Plastic applicators have the least radiation attenuation compared to the other applicators, therefore, they are recommended for use in brachytherapy. A table of correction factors has been introduced for different sources and applicators with different materials for the clinical applications.

Original languageEnglish
Pages (from-to)477-491
Number of pages15
JournalAustralasian Physical and Engineering Sciences in Medicine
Volume39
Issue number2
DOIs
StatePublished - Jun 1 2016

Bibliographical note

Publisher Copyright:
© 2016, Australasian College of Physical Scientists and Engineers in Medicine.

Keywords

  • Applicator
  • Brachytherapy
  • Dose distribution
  • Dosimetric parameters
  • Monte Carlo simulation

ASJC Scopus subject areas

  • Biophysics
  • Biomedical Engineering
  • Radiology Nuclear Medicine and imaging
  • General Physics and Astronomy

Fingerprint

Dive into the research topics of 'Evaluating the effect of various intracavitary applicators on dosimetric parameters of 192Ir, 137Cs, and 60Co sources'. Together they form a unique fingerprint.

Cite this