Abstract
Purpose: To quantify the differences in dosimetry as a function of ipsilateral lung density and treatment delivery parameters for stereotactic, single dose of volumetric modulated arc therapy (VMAT) lung stereotactic body radiation therapy (SBRT) delivered with 6X flattening filter free (6X-FFF) beams compared to traditional flattened 6X (6X-FF) beams. Materials/methods: Thirteen consecutive early stage I–II non-small-cell-lung cancer (NSCLC) patients were treated with highly conformal noncoplanar VMAT SBRT plans (3–6 partial arcs) using 6X-FFF beam and advanced Acuros-based dose calculations to a prescription dose of 30 Gy in one fraction to the tumor margin. These clinical cases included relatively smaller tumor (island tumors) sizes (2.0–4.2 cm diameters) and varying average ipsilateral lung densities between 0.14 g/cc and 0.34 g/cc. Treatment plans were reoptimized with 6X-FF beams for identical beam/arc geometries and planning objectives. For same target coverage, the organs-at-risk (OAR) dose metrics as a function of ipsilateral lung density were compared between 6X-FFF and 6X-FF plans. Moreover, monitor units (MU), beam modulation factor (MF) and beam-on time (BOT) were evaluated. Results: Both plans met the RTOG-0915 protocol compliance. The ipsilateral lung density and the tumor location heavily influenced the treatment plans with 6X-FFF and 6X-FF beams, showing differences up to 12% for the gradient indices. For similar target coverage, 6X-FFF beams showed better target conformity, lower intermediate dose-spillage, and lower dose to the OAR. Additionally, BOT was reduced by a factor of 2.3 with 6X-FFF beams compared to 6X-FF beams. Conclusion: While prescribing dose to the tumor periphery, 6X-FFF VMAT plans for stereotactic single-dose lung SBRT provided similar target coverage with better dose conformity, superior intermediate dose-spillage (improved dose coverage at tumor interface), and improved OAR sparing compared to traditional 6X-FF beams and significantly reduced treatment time. The ipsilateral lung density and tumor location considerably affected dose distributions requiring special attention for clinical SBRT plan optimization on a per-patient basis. Clinical follow up of these patients for tumor local-control rate and treatment-related toxicities is in progress.
| Original language | English |
|---|---|
| Pages (from-to) | 26-35 |
| Number of pages | 10 |
| Journal | Journal of Applied Clinical Medical Physics |
| Volume | 21 |
| Issue number | 1 |
| DOIs | |
| State | Published - Jan 1 2020 |
Bibliographical note
Publisher Copyright:© 2019 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.
UN SDGs
This output contributes to the following UN Sustainable Development Goals (SDGs)
-
SDG 3 Good Health and Well-being
Keywords
- FFF-beam
- RTOG-0915
- lung SBRT
- noncoplanar VMAT
- single-dose
ASJC Scopus subject areas
- Radiation
- Instrumentation
- Radiology Nuclear Medicine and imaging
Fingerprint
Dive into the research topics of 'FFF-VMAT for SBRT of lung lesions: Improves dose coverage at tumor-lung interface compared to flattened beams'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver