Insertion of Bicaval Dual Lumen extracorporeal membrane oxygenation catheter with image guidance

Jeffrey Javidfar, Dongfang Wang, Joseph B. Zwischenberger, Joseph Costa, Linda Mongero, Joshua Sonett, Matthew Bacchetta

Research output: Contribution to journalArticlepeer-review

106 Scopus citations

Abstract

Correct placement of the Avalon Elite Bicaval Dual Lumen catheter (Avalon Laboratories, LLC, CA) for single-site venovenous extracorporeal membrane oxygenation (VV ECMO) is safe using image guidance. Using this technique, 26 of 27 patients (96%) had uneventful placement of the cannula in the right internal jugular vein. One patient had a superior vena cava injury during serial dilation, and another patient required cannula repositioning for improved flows. We recommend using both fluoroscopy and transesophageal echocardiogram (TEE) for training purposes or during initial use of the Avalon Elite. As proficiency improves, TEE at the bedside provides an excellent standard of care. Double-lumen ECMO catheters can be effectively placed under image guidance with minimal need for repositioning.

Original languageEnglish
Pages (from-to)203-205
Number of pages3
JournalASAIO Journal
Volume57
Issue number3
DOIs
StatePublished - May 2011

ASJC Scopus subject areas

  • Biophysics
  • Bioengineering
  • Biomaterials
  • Biomedical Engineering

Fingerprint

Dive into the research topics of 'Insertion of Bicaval Dual Lumen extracorporeal membrane oxygenation catheter with image guidance'. Together they form a unique fingerprint.

Cite this