T1 - Multiple autoclave cycle effects on cyclic fatigue of nickel-titanium rotary files produced by new manufacturing methods
AU - Hilfer, Paul B.
AU - Bergeron, Brian E.
AU - Mayerchak, Michael J.
AU - Roberts, Howard W.
AU - Jeansonne, Billie G.
N1 - Funding Information:
Supported by the AAE Foundation .
PY - 2011/1
Y1 - 2011/1
N2 - Introduction: Novel nickel-titanium rotary files with proprietary manufacturing techniques have recently been marketed. The purpose of this study was to assess multiple autoclave cycle effects on cyclic fatigue of GT Series X files (Dentsply Tulsa Dental Specialties, Tulsa, OK) and Twisted Files (SybronEndo, Orange, CA) Methods: A jig using a 5-mm radius curve with 90° of maximum file flexure was used to induce cyclic fatigue failure. Files (n = 10) representing each experimental group (GT Series X 20/.04 and 20/.06; Twisted Files 25/.04 and 25/.06) were first tested to establish baseline mean cycles to failure (MCF). Experimental groups (n = 20) were then cycled to 25% of the established baseline MCF and then autoclaved. Additional autoclaving was accomplished at 50% and 75% of MCF followed by continual testing until failure. Control groups (n = 20) underwent the same procedures except autoclaving was not accomplished. Results: The GT Series X (20/.04 and 20/.06) files showed no significant difference (p = 0.918/p = 0.096) in MCF for experimental versus control files. Twisted Files (25/.04) showed no significant difference (p = 0.432) in MCF between experimental and control groups. However, the Twisted Files (25/.06) experimental group showed a significantly lower (p = 0.0175) MCF compared with the controls. Conclusions: Under the conditions of this evaluation, autoclave sterilization significantly decreased cyclic fatigue resistance of one of the four file groups tested. Repeated autoclaving significantly reduced the MCF of 25/.06 Twisted Files; however, 25/.04 Twisted Files and both GT Series X files tested were not significantly affected by the same conditions.
AB - Introduction: Novel nickel-titanium rotary files with proprietary manufacturing techniques have recently been marketed. The purpose of this study was to assess multiple autoclave cycle effects on cyclic fatigue of GT Series X files (Dentsply Tulsa Dental Specialties, Tulsa, OK) and Twisted Files (SybronEndo, Orange, CA) Methods: A jig using a 5-mm radius curve with 90° of maximum file flexure was used to induce cyclic fatigue failure. Files (n = 10) representing each experimental group (GT Series X 20/.04 and 20/.06; Twisted Files 25/.04 and 25/.06) were first tested to establish baseline mean cycles to failure (MCF). Experimental groups (n = 20) were then cycled to 25% of the established baseline MCF and then autoclaved. Additional autoclaving was accomplished at 50% and 75% of MCF followed by continual testing until failure. Control groups (n = 20) underwent the same procedures except autoclaving was not accomplished. Results: The GT Series X (20/.04 and 20/.06) files showed no significant difference (p = 0.918/p = 0.096) in MCF for experimental versus control files. Twisted Files (25/.04) showed no significant difference (p = 0.432) in MCF between experimental and control groups. However, the Twisted Files (25/.06) experimental group showed a significantly lower (p = 0.0175) MCF compared with the controls. Conclusions: Under the conditions of this evaluation, autoclave sterilization significantly decreased cyclic fatigue resistance of one of the four file groups tested. Repeated autoclaving significantly reduced the MCF of 25/.06 Twisted Files; however, 25/.04 Twisted Files and both GT Series X files tested were not significantly affected by the same conditions.
KW - Autoclave
KW - cyclic fatigue
KW - nickel-titanium
UR - http://www.scopus.com/inward/record.url?scp=78650085817&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=78650085817&partnerID=8YFLogxK
U2 - 10.1016/j.joen.2010.09.011
DO - 10.1016/j.joen.2010.09.011
M3 - Article
C2 - 21146081
AN - SCOPUS:78650085817
SN - 0099-2399
VL - 37
SP - 72
EP - 74
JO - Journal of Endodontics
JF - Journal of Endodontics
IS - 1
ER -