Multiple failed external defibrillation attempts during robot-assisted internal mammary harvest for myocardial revascularization

Kevin W. Hatton, Lawrence C. Kilinski, Chandrashekar Ramaiah, Randall M. Schell

Research output: Contribution to journalArticlepeer-review

22 Scopus citations

Abstract

We describe multiple failed external defibrillation attempts via rescue defibrillation pads for ventricular fibrillation that occurred secondary to direct electrical current transmission through the pericardial sac from electrocautery during robot-assisted internal mammary harvest. Only after resumption of two-lung ventilation and decompression of the iatrogenic pneumothorax was the patient successfully defibrillated. Conditions necessary for robotic intrathoracic surgery may make defibrillation and resuscitation difficult if they become necessary.

Original languageEnglish
Pages (from-to)1113-1114
Number of pages2
JournalAnesthesia and Analgesia
Volume103
Issue number5
DOIs
StatePublished - Nov 2006

Funding

Supported by Department of Anesthesiology, University of Kentucky Chandler Medical Center.

FundersFunder number
University of Kentucky Chandler Medical Center, Lexington, KY

    ASJC Scopus subject areas

    • Anesthesiology and Pain Medicine

    Fingerprint

    Dive into the research topics of 'Multiple failed external defibrillation attempts during robot-assisted internal mammary harvest for myocardial revascularization'. Together they form a unique fingerprint.

    Cite this