TY - JOUR
T1 - Significant reduction in circuit pressure with modified plasma separation chamber for a heparin removal device
AU - Tao, Weike
AU - Deyo, Donald J.
AU - Alpard, Scott K.
AU - Vertrees, Roger A.
AU - Hoang, Vivi
AU - Zwischenberger, Joseph B.
PY - 1999
Y1 - 1999
N2 - The heparin removal device (HRD), using plasma separation and poly-L-lysine (PLL) affinity adsorption is an effective alternative to protamine after cardiopulmonary bypass. Previous designs of the HRD used standard port connections between the extracorporeal tubing and the plasma separation chambers, which resulted in excessively high circuit pressures. To reduce circuit pressure, the connection ports are enlarged, keeping other circuit components and sorbent amounts unchanged.
AB - The heparin removal device (HRD), using plasma separation and poly-L-lysine (PLL) affinity adsorption is an effective alternative to protamine after cardiopulmonary bypass. Previous designs of the HRD used standard port connections between the extracorporeal tubing and the plasma separation chambers, which resulted in excessively high circuit pressures. To reduce circuit pressure, the connection ports are enlarged, keeping other circuit components and sorbent amounts unchanged.
UR - http://www.scopus.com/inward/record.url?scp=0032601381&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0032601381&partnerID=8YFLogxK
U2 - 10.1097/00002480-199901000-00012
DO - 10.1097/00002480-199901000-00012
M3 - Article
C2 - 9952007
AN - SCOPUS:0032601381
SN - 1058-2916
VL - 45
SP - 50
EP - 52
JO - ASAIO Journal
JF - ASAIO Journal
IS - 1