TY - JOUR
T1 - T 1 mapping performance and measurement repeatability
T2 - Results from the multi-national T 1 mapping standardization phantom program (T1MES)
AU - Captur, Gabriella
AU - Bhandari, Abhiyan
AU - Brühl, Rüdiger
AU - Ittermann, Bernd
AU - Keenan, Kathryn E.
AU - Yang, Ye
AU - Eames, Richard J.
AU - Benedetti, Giulia
AU - Torlasco, Camilla
AU - Ricketts, Lewis
AU - Boubertakh, Redha
AU - Fatih, Nasri
AU - Greenwood, John P.
AU - Paulis, Leonie E.M.
AU - Lawton, Chris B.
AU - Bucciarelli-Ducci, Chiara
AU - Lamb, Hildo J.
AU - Steeds, Richard
AU - Leung, Steve W.
AU - Berry, Colin
AU - Valentin, Sinitsyn
AU - Flett, Andrew
AU - De Lange, Charlotte
AU - Decobelli, Francesco
AU - Viallon, Magalie
AU - Croisille, Pierre
AU - Higgins, David M.
AU - Greiser, Andreas
AU - Pang, Wenjie
AU - Hamilton-Craig, Christian
AU - Strugnell, Wendy E.
AU - Dresselaers, Tom
AU - Barison, Andrea
AU - Dawson, Dana
AU - Taylor, Andrew J.
AU - Mongeon, François Pierre
AU - Plein, Sven
AU - Messroghli, Daniel
AU - Al-Mallah, Mouaz
AU - Grieve, Stuart M.
AU - Lombardi, Massimo
AU - Jang, Jihye
AU - Salerno, Michael
AU - Chaturvedi, Nish
AU - Kellman, Peter
AU - Bluemke, David A.
AU - Nezafat, Reza
AU - Gatehouse, Peter
AU - Moon, James C.
N1 - Publisher Copyright:
© 2020 The Author(s).
PY - 2020/5/7
Y1 - 2020/5/7
N2 - Background: The T 1 Mapping and Extracellular volume (ECV) Standardization (T1MES) program explored T 1 mapping quality assurance using a purpose-developed phantom with Food and Drug Administration (FDA) and Conformité Européenne (CE) regulatory clearance. We report T 1 measurement repeatability across centers describing sequence, magnet, and vendor performance. Methods: Phantoms batch-manufactured in August 2015 underwent 2 years of structural imaging, B 0 and B 1, and "reference" slow T 1 testing. Temperature dependency was evaluated by the United States National Institute of Standards and Technology and by the German Physikalisch-Technische Bundesanstalt. Center-specific T 1 mapping repeatability (maximum one scan per week to minimum one per quarter year) was assessed over mean 358 (maximum 1161) days on 34 1.5 T and 22 3 T magnets using multiple T 1 mapping sequences. Image and temperature data were analyzed semi-automatically. Repeatability of serial T 1 was evaluated in terms of coefficient of variation (CoV), and linear mixed models were constructed to study the interplay of some of the known sources of T 1 variation. Results: Over 2 years, phantom gel integrity remained intact (no rips/tears), B 0 and B 1 homogenous, and "reference" T 1 stable compared to baseline (% change at 1.5 T, 1.95 ± 1.39%; 3 T, 2.22 ± 1.44%). Per degrees Celsius, 1.5 T, T 1 (MOLLI 5s(3s)3s) increased by 11.4 ms in long native blood tubes and decreased by 1.2 ms in short post-contrast myocardium tubes. Agreement of estimated T 1 times with "reference" T 1 was similar across Siemens and Philips CMR systems at both field strengths (adjusted R 2 ranges for both field strengths, 0.99-1.00). Over 1 year, many 1.5 T and 3 T sequences/magnets were repeatable with mean CoVs < 1 and 2% respectively. Repeatability was narrower for 1.5 T over 3 T. Within T1MES repeatability for native T 1 was narrow for several sequences, for example, at 1.5 T, Siemens MOLLI 5s(3s)3s prototype number 448B (mean CoV = 0.27%) and Philips modified Look-Locker inversion recovery (MOLLI) 3s(3s)5s (CoV 0.54%), and at 3 T, Philips MOLLI 3b(3s)5b (CoV 0.33%) and Siemens shortened MOLLI (ShMOLLI) prototype 780C (CoV 0.69%). After adjusting for temperature and field strength, it was found that the T 1 mapping sequence and scanner software version (both P < 0.001 at 1.5 T and 3 T), and to a lesser extent the scanner model (P = 0.011, 1.5 T only), had the greatest influence on T 1 across multiple centers. Conclusion: The T1MES CE/FDA approved phantom is a robust quality assurance device. In a multi-center setting, T 1 mapping had performance differences between field strengths, sequences, scanner software versions, and manufacturers. However, several specific combinations of field strength, sequence, and scanner are highly repeatable, and thus, have potential to provide standardized assessment of T 1 times for clinical use, although temperature correction is required for native T 1 tubes at least.
AB - Background: The T 1 Mapping and Extracellular volume (ECV) Standardization (T1MES) program explored T 1 mapping quality assurance using a purpose-developed phantom with Food and Drug Administration (FDA) and Conformité Européenne (CE) regulatory clearance. We report T 1 measurement repeatability across centers describing sequence, magnet, and vendor performance. Methods: Phantoms batch-manufactured in August 2015 underwent 2 years of structural imaging, B 0 and B 1, and "reference" slow T 1 testing. Temperature dependency was evaluated by the United States National Institute of Standards and Technology and by the German Physikalisch-Technische Bundesanstalt. Center-specific T 1 mapping repeatability (maximum one scan per week to minimum one per quarter year) was assessed over mean 358 (maximum 1161) days on 34 1.5 T and 22 3 T magnets using multiple T 1 mapping sequences. Image and temperature data were analyzed semi-automatically. Repeatability of serial T 1 was evaluated in terms of coefficient of variation (CoV), and linear mixed models were constructed to study the interplay of some of the known sources of T 1 variation. Results: Over 2 years, phantom gel integrity remained intact (no rips/tears), B 0 and B 1 homogenous, and "reference" T 1 stable compared to baseline (% change at 1.5 T, 1.95 ± 1.39%; 3 T, 2.22 ± 1.44%). Per degrees Celsius, 1.5 T, T 1 (MOLLI 5s(3s)3s) increased by 11.4 ms in long native blood tubes and decreased by 1.2 ms in short post-contrast myocardium tubes. Agreement of estimated T 1 times with "reference" T 1 was similar across Siemens and Philips CMR systems at both field strengths (adjusted R 2 ranges for both field strengths, 0.99-1.00). Over 1 year, many 1.5 T and 3 T sequences/magnets were repeatable with mean CoVs < 1 and 2% respectively. Repeatability was narrower for 1.5 T over 3 T. Within T1MES repeatability for native T 1 was narrow for several sequences, for example, at 1.5 T, Siemens MOLLI 5s(3s)3s prototype number 448B (mean CoV = 0.27%) and Philips modified Look-Locker inversion recovery (MOLLI) 3s(3s)5s (CoV 0.54%), and at 3 T, Philips MOLLI 3b(3s)5b (CoV 0.33%) and Siemens shortened MOLLI (ShMOLLI) prototype 780C (CoV 0.69%). After adjusting for temperature and field strength, it was found that the T 1 mapping sequence and scanner software version (both P < 0.001 at 1.5 T and 3 T), and to a lesser extent the scanner model (P = 0.011, 1.5 T only), had the greatest influence on T 1 across multiple centers. Conclusion: The T1MES CE/FDA approved phantom is a robust quality assurance device. In a multi-center setting, T 1 mapping had performance differences between field strengths, sequences, scanner software versions, and manufacturers. However, several specific combinations of field strength, sequence, and scanner are highly repeatable, and thus, have potential to provide standardized assessment of T 1 times for clinical use, although temperature correction is required for native T 1 tubes at least.
KW - Calibration
KW - Extracellular volume
KW - Phantom
KW - Repeatability
KW - Standardization
KW - T mapping
UR - http://www.scopus.com/inward/record.url?scp=85084329171&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85084329171&partnerID=8YFLogxK
U2 - 10.1186/s12968-020-00613-3
DO - 10.1186/s12968-020-00613-3
M3 - Article
C2 - 32375896
AN - SCOPUS:85084329171
SN - 1097-6647
VL - 22
JO - Journal of Cardiovascular Magnetic Resonance
JF - Journal of Cardiovascular Magnetic Resonance
IS - 1
M1 - 31
ER -