Abstract
A completely implantable, one piece electromechanical total artificial heart (TAH) intended for permanent human use was developed. It consisted of left and right conically shaped pusher-plate blood pumps sandwiching a thin centerpiece with a compact, efficient electromechanical actuator. The actuator consisted of a direct current brushless motor; a planetary roller screw fit the space between the two conically shaped pusher-plates. The rotational motion of the motor was converted to the rectilinear motion of the rollerscrew to displace the left and right pusher-plates in the left master alternate mode. The diameter of the assembled TAH was 97 mm, with a central thickness of 82 mm. The overall weight was 620 g, with a displaced volume of 510 ml. The pump provided flows of 3-8 L/min with a preload of 1-15 mmHg against an afterload of 100 mmHg. The net efficiency ranged from 15% to 18%. This model showed good fit in the pericardial space of heart transplant recipients (body weight, 77 kg).
| Original language | English |
|---|---|
| Pages (from-to) | M238-M240 |
| Journal | ASAIO Transactions |
| Volume | 37 |
| Issue number | 3 |
| State | Published - Jul 1991 |