A comparison of solid phantoms with water for dosimetry of 1 2 5I brachytherapy sources

Producción científica: Articlerevisión exhaustiva

111 Citas (Scopus)

Resumen

Dosimetry of brachytherapy sources is critically dependent on precise measurement of the source–detector distance. A solid phantom can be precisely machined and hence distances can be accurately determined. In this work LiF thermoluminescent chips are used for absolute dose rate measurements in solid water, polymethylmethacrylate (PMMA), and polystyrene. These media are examined for their suitability in the dosimetry of 1 2 5I by comparing depth doses in each phantom. Measurements and Monte Carlo calculations show that solid water is equivalent to water for the dosimetry of 1 2 5I seeds, however, polystyrene and PMMA are not equivalent to water. Also, photon energy spectra for several depths in each phantom material have been calculated and are used to determine average photon energy and mass energy absorption coefficients as a function of depth.

Idioma originalEnglish
Páginas (desde-hasta)695-701
Número de páginas7
PublicaciónMedical Physics
Volumen15
N.º5
DOI
EstadoPublished - sept 1988

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging