A novel less invasive single port pulsatile ventricular assist device

Producción científica: Articlerevisión exhaustiva

Resumen

A less invasive paracorporeal single port pulsatile ventricular assist device (sppVAD) was developed for cardiogenic shock. This compact sppVAD includes a valved single lumen cannula (VSLC), valveless single port diaphragm displacement pump (spDDP), and portable console. A 33/27 Fr VSLC was fabricated for mock loop testing, a 28/25 Fr VSLC was made for 6 hours sheep testing (n = 5, 44–52 kg), and valveless spDDP with portable console was made. The sppVAD (33/27 Fr VSLC) achieved 6.2 L/min flow in mock loop. In sheep, the 28/25 Fr VSLC was inserted without cardiopulmonary bypass from apex to ascending aorta within 1 minute after apex exposure. The sppVAD achieved 3.75 L/min flow in sheep with stable hemodynamics. Hemoglobin, free hemoglobin, and platelet/white blood cell counts were unchanged through 6 hours. Thus, our less invasive sppVAD achieved total left ventricle support with acceptable short-term hemocompatibility, showing good promise as a temporary VAD for cardiogenic shock.

Idioma originalEnglish
Número de artículo100540
PublicaciónJHLT Open
Volumen12
DOI
EstadoPublished - may 2026

Nota bibliográfica

Publisher Copyright:
© 2026 The Author(s)

Financiación

This work was supported by an NIH Phase I SBIR grant (R43HL158430).

FinanciadoresNúmero del financiador
National Institutes of Health (NIH)
Small Business Innovation ResearchR43HL158430

    ASJC Scopus subject areas

    • Medicine (miscellaneous)
    • Pulmonary and Respiratory Medicine
    • Cardiology and Cardiovascular Medicine

    Huella

    Profundice en los temas de investigación de 'A novel less invasive single port pulsatile ventricular assist device'. En conjunto forman una huella única.

    Citar esto