Accuracy of pulsatile photoplethysmography applications or handheld devices vs. 12-lead ECG for atrial fibrillation screening: a systematic review and meta-analysis

  • Yasar Sattar
  • , David Song
  • , Deepika Sarvepalli
  • , Syeda Ramsha Zaidi
  • , Waqas Ullah
  • , Junaid Arshad
  • , Tanveer Mir
  • , Mohamed Zghouzi
  • , Islam Y. Elgendy
  • , Waqas Qureshi
  • , Nagib Chalfoun
  • , MChadi C. Alraies

Producción científica: Articlerevisión exhaustiva

8 Citas (Scopus)

Resumen

Background: The relative accuracy of pulsatile photoplethysmography applications (PPG) or handheld (HH) devices compared with the gold standard 12-lead electrocardiogram (ECG) for the diagnosis of atrial fibrillation is unknown. Methods: Digital databases were searched to identify relevant articles. Raw data were pooled using a bivariate model to calculate diagnostic accuracy measures and estimate Hierarchical Summary Receiver Operating Characteristic (HSROC). Results: A total of 10 articles comprising 4296 patients (mean age 68.9 years, with 56% males) were included in the analysis. Compared with EKG, the pooled sensitivity of PPG and HH devices in AF detection was 0.93 (95% CI 0.87–0.96; p < 0.05) and 0.87 (95% CI. 0.74–0.94; p < 0.05), respectively. The pooled specificity of PPG and HH devices in AF detection was 0.91 (95% CI 0.88–0.94; p < 0.05) and 0.96 (95% CI 0.90–0.98; p < 0.05), respectively. The diagnostic odds ratio was 129 and 144 for PPG and HH devices, respectively. Fagan’s nomogram showed the probability of a patient having AF and normal rhythm on PPG or HH devices was 2–3%, while the post-test probability of having AF with an irregular R-R interval on PPG or HH devices was 73% and 82%, respectively. The scatter plot of positive and negative likelihood ratio showed high confirmation of AF and reliability of exclusion of absence of irregular R-R intervals (positive likelihood ratio > 10, and negative likelihood ratio < 0.1) on HH devices while PPG was used as confirmation only. Conclusions: The PPG or HH devices can serve as a reliable alternative for the detection of AF.

Idioma originalEnglish
Páginas (desde-hasta)33-44
Número de páginas12
PublicaciónJournal of Interventional Cardiac Electrophysiology
Volumen65
N.º1
DOI
EstadoPublished - oct 2022

Nota bibliográfica

Publisher Copyright:
© 2021, Springer Science+Business Media, LLC, part of Springer Nature.

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine
  • Physiology (medical)

Huella

Profundice en los temas de investigación de 'Accuracy of pulsatile photoplethysmography applications or handheld devices vs. 12-lead ECG for atrial fibrillation screening: a systematic review and meta-analysis'. En conjunto forman una huella única.

Citar esto