An investigation into the feasibility and efficacy of stereotactic radiosurgery for 1–3 cm single brain lesions on the ring-mounted Halcyon LINAC

  • Kate Hazelwood
  • , Shane McCarthy
  • , Josh Misa
  • , David Castelvetere
  • , William St. Clair
  • , Damodar Pokhrel

Producción científica: Articlerevisión exhaustiva

3 Citas (Scopus)

Resumen

Purpose: An evaluation of the accuracy, safety, and efficiency of the Halcyon ring delivery system (RDS) for stereotactic radiosurgery (SRS) treatment to relatively small (1–3 cm) brain lesions. Methods: After completing the extensive in-house quality assurance checks including Winston–Lutz test and independent dose verification via MD Anderson IROC SRS head phantom irradiation on Halcyon, fifteen brain SRS patients previously treated with a single dose of 20 Gy on TrueBeam (6MV-FFF) with HyperArc geometry were retrospectively replanned on Halcyon (6MV-FFF). Plan quality metrics including conformity index (CI), gradient index (GI), gradient distance (GD), PTV coverage, gross tumor volume (GTV) dose, heterogeneity index (HI), and doses to organs-at-risk (OAR) including normal brain dose were evaluated. Patient-specific quality assurance (PSQA) and independent dose verification via in-house Monte Carlo (MC) 2nd checks were performed. Results: The Halcyon was able to provide highly conformal brain SRS plans. When compared to TrueBeam, CI, planning target volume (PTV) coverage, GTV dose (mean and minimum), HI, and doses to brainstem, optic pathway, and cochlea were statistically insignificant. Statistically significant increases in GI (3.76 vs. 3.25, p < 0.001), GD (0.56 cm vs. 0.48 cm, p = 0.001), and V12Gy (5.5 cc vs. 4.6 cc, p = 0.014), on average using Halcyon versus TrueBeam was found, albeit clinically acceptable values for the majority of brain SRS cases. Halcyon plans provided statistically insignificant maximum dose to most adjacent OARs, though there was a statistically significant decrease in the maximum dose to the spinal cord (0.1 Gy vs. 0.4 Gy, p = 0.009). Halcyon beam-on time increases by a factor of ∼2 (p < 0.001). However, the faster patient setup on Halcyon results in a comparable estimated overall treatment time for both platforms. Plan deliverability and accuracy was ensured with PSQA (> 95% pass rate for 2%/2 mm clinical gamma criteria) results and MC 2nd check agreement within ± 5.0%. Conclusions: Halcyon brain SRS plans provided a similar plan quality compared to HyperArc plans, although it demonstrated an inferior intermediate dose fall off thus slightly higher V12Gy. This study suggests that Halcyon provides acceptable treatment for solitary relatively small brain lesions of 1–3 cm in diameter. Treatment of select patients on Halcyon will be started at our clinic and it is recommended that other clinics complete an end-to-end test, validate, and implement Halcyon SRS treatments at their practices, especially community cancer centers to provide high-quality service to an underserved patient cohort.

Idioma originalEnglish
Número de artículoe70105
PublicaciónJournal of Applied Clinical Medical Physics
Volumen26
N.º6
DOI
EstadoPublished - jun 2025

Nota bibliográfica

Publisher Copyright:
© 2025 The Author(s). Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine.

Financiación

The authors would like to express their sincere gratitude to all anonymous reviewers for their time and constructive feedback that allowed us to improve the clarity of this technical report. Authors would also like to express their sincere thanks to all faculty & staff for their support and help to the AAPM 2024 Summer Undergraduate Fellow (Department of Radiation Medicine, University of Kentucky, Lexington, KY). Moreover, authors would like to thank the MD Anderson's IROC Houston QA Center (Houston, TX) for providing the SRS head phantom for credentialling our Halcyon LINAC for brain SRS/SRT treatments. This work was partially supported by the AAPM 2024 Summer Undergraduate Fellowship Program and the Varian Medical Systems (Palo Alto, CA).

Financiadores
University of Texas Anderson Cancer Center
Scoliosis Research Society Research Committee
American Association of Physicists in Medicine
Varian Medical Systems

    ODS de las Naciones Unidas

    Este resultado contribuye a los siguientes Objetivos de Desarrollo Sostenible

    1. Good health and well being
      Good health and well being

    ASJC Scopus subject areas

    • Radiation
    • Instrumentation
    • Radiology Nuclear Medicine and imaging

    Huella

    Profundice en los temas de investigación de 'An investigation into the feasibility and efficacy of stereotactic radiosurgery for 1–3 cm single brain lesions on the ring-mounted Halcyon LINAC'. En conjunto forman una huella única.

    Citar esto