Application of a color scanner for 60Co high dose rate brachytherapy dosimetry with EBT radiochromic film

  • Mahdi Ghorbani
  • , Mohammad Taghi Bahreyni Toossi
  • , Ali Asghar Mowlavi
  • , Shahram Bayani Roodi
  • , Ali Soleimani Meigooni

Producción científica: Articlerevisión exhaustiva

4 Citas (Scopus)

Resumen

Background. The aim of this study is to evaluate the performance of a color scanner as a radiochromic film reader in two dimensional dosimetry around a high dose rate brachytherapy source. Materials and methods. A Microtek ScanMaker 1000XL film scanner was utilized for the measurement of dose distribution around a high dose rate GZP6 60Co brachytherapy source with GafChromic® EBT radiochromic films. In these investigations, the non-uniformity of the film and scanner response, combined, as well as the films sensitivity to scanner's light source was evaluated using multiple samples of films, prior to the source dosimetry. The results of these measurements were compared with the Monte Carlo simulated data using MCNPX code. In addition, isodose curves acquired by radiochromic films and Monte Carlo simulation were compared with those provided by the GZP6 treatment planning system. Results. Scanning of samples of uniformly irradiated films demonstrated approximately 2.85% and 4.97% nonuniformity of the response, respectively in the longitudinal and transverse directions of the film. Our findings have also indicated that the film response is not affected by the exposure to the scanner's light source, particularly in multiple scanning of film. The results of radiochromic film measurements are in good agreement with the Monte Carlo calculations (4%) and the corresponding dose values presented by the GZP6 treatment planning system (5%). Conclusions. The results of these investigations indicate that the Microtek ScanMaker 1000XL color scanner in conjunction with GafChromic EBT film is a reliable system for dosimetric evaluation of a high dose rate brachytherapy source.

Idioma originalEnglish
Páginas (desde-hasta)363-369
Número de páginas7
PublicaciónRadiology and Oncology
Volumen46
N.º4
DOI
EstadoPublished - dic 2012

Financiación

The authors would like to thank Dr. Abdol-Reza Hashemian for his assistance by providing the radiochromic film sheets. We also would like to acknowledge Dr. Ghorban Safaeian for film calibrations. This work was funded by office of vice president for research of Mashhad University of Medical Sciences.

Financiadores
Mashhad University of Medical Sciences

    ASJC Scopus subject areas

    • Oncology
    • Radiology Nuclear Medicine and imaging

    Huella

    Profundice en los temas de investigación de 'Application of a color scanner for 60Co high dose rate brachytherapy dosimetry with EBT radiochromic film'. En conjunto forman una huella única.

    Citar esto