Ir directamente a la navegación principal Ir directamente a la búsqueda Ir directamente al contenido principal

Aseptic loosening of an APTIS distal radioulnar joint arthroplasty – Case report

  • Karis Bennett
  • , Srinath Kamineni

Producción científica: Articlerevisión exhaustiva

4 Citas (Scopus)

Resumen

Dysfunction of the distal radioulnar joint can cause significant pain and instability. The self-stabilizing APTIS distal radioulnar joint prosthesis is used as a solution for severe distal radioulnar joint pathologies. We present a case of a 60-year-old male, who received an APTIS distal radioulnar joint prosthesis which resulted in aseptic loosening within five years of the initial implantation. Infection, incorrect implantation, demographic differences and over-activity were all excluded as the source; therefore, mechanical aseptic loosening was concluded. Ultimately, two surgeries were required to resolve the patient's pain, which resulted in a one-bone forearm once the implant was extracted. The solution to a failed APTIS implant, a one bone forearm, is difficult and protracted, so every effort should be attempted to preserve distal ulna bone stock before resorting to the implantation of this device.

Idioma originalEnglish
Número de artículo100196
PublicaciónTrauma Case Reports
Volumen21
DOI
EstadoPublished - jun 2019

Nota bibliográfica

Publisher Copyright:
© 2019

ASJC Scopus subject areas

  • Emergency Medicine
  • Orthopedics and Sports Medicine
  • Critical Care and Intensive Care Medicine

Huella

Profundice en los temas de investigación de 'Aseptic loosening of an APTIS distal radioulnar joint arthroplasty – Case report'. En conjunto forman una huella única.

Citar esto