Assessment of the feasibility of determination of cholesterol and other blood constituents by near-infrared reflectance analysis

  • Robert A. Lodder
  • , Gary M. Hieftje
  • , Wells Moorehead
  • , Steven P. Robertson
  • , Phillip Rand

Producción científica: Article › revisión exhaustiva

12 Citas (Scopus)

Resumen

Near-infrared reflectance spectrometry of blood serum can yield values for serum cholesterol that correlate reasonably well (r = 0.96) with those from common reference analytical methods. However, the variability of serum can cause ostensibly validated calibrations to fail on new samples. The determination of blood components such as cholesterol and triglycerides by near-infrared reflectance is complicated by their low concentrations, the variety of forms in which they appear, and by the natural variability of the blood matrix. These difficulties, when combined with the problems encountered in obtaining a representative sample from a given individual, can make it almost impossible to select, by a regression procedure, a wavelength combination that is characteristic of the complete blood matrix. The failure of the regression process to find characteristic wavelengths generates a false-sample problem in which even small changes at the analytical wavelengths produce a grossly unreliable cholesterol or triglyceride determination.

Idioma originalEnglish
Páginas (desde-hasta)193-198
Número de páginas6
PublicaciónTalanta
Volumen36
N.º1-2
DOI
EstadoPublished - 1989

Nota bibliográfica

Funding Information:
Acknowledgements-Thisw ork was supportedi n part by the National Science Foundation through Grant CHE 87-22639b,y the Office of Naval Research,a nd by Miles LaboratorieIsn, c.

Financiación

Acknowledgements-Thisw ork was supportedi n part by the National Science Foundation through Grant CHE 87-22639b,y the Office of Naval Research,a nd by Miles LaboratorieIsn, c.

FinanciadoresNúmero del financiador
Miles LaboratorieIsn
National Science Foundation (NSF)CHE 87-22639b
Office of Naval Research

    ASJC Scopus subject areas

    • Analytical Chemistry

    Huella

    Profundice en los temas de investigación de 'Assessment of the feasibility of determination of cholesterol and other blood constituents by near-infrared reflectance analysis'. En conjunto forman una huella única.

    Citar esto