Ir directamente a la navegación principal Ir directamente a la búsqueda Ir directamente al contenido principal

Calibration and field deployment of low-cost fluid flow-rate sensors using a wireless network

  • Cindy K. Harnett
  • , Matthew T. Schueler
  • , Nick R. Blumenthal
  • , Kristy L. Hopf
  • , Jimmy F. Fox
  • , Sruti Pulugurtha

Producción científica: Article › revisión exhaustiva

9 Citas (Scopus)

Resumen

Low-cost networked fluid flow velocity sensors are needed for high-density sampling in environmental research and other applications requiring automated fluid flow velocity mapping. The flow sensor in this paper is the target type consisting of a flap that deflects in the flow, changing the resistance of a strain gauge. A highly compliant flexible resistor promises to reduce costs by serving as the flow target and by eliminating the signal amplification circuitry needed for conventional strain gauges. However, we show that the individual calibration of these resistive sensors is critical for recovering the stream flow velocity. A wireless network is used for the mass calibration of multiple sensors in a test flume. After individual sensor calibrations are obtained, the same wireless network can be used to collect flow velocity data in the sensor application. A modular connection system enables the user to reconfigure quickly the system's physical layout for calibration or deployment purposes. The results are reported on an outdoor deployment of the flow sensors for logging stream flow data in environmental research.

Idioma originalEnglish
Número de artículo5491170
Páginas (desde-hasta)633-641
Número de páginas9
PublicaciónIEEE Transactions on Instrumentation and Measurement
Volumen60
N.º2
DOI
EstadoPublished - feb 2011

Nota bibliográfica

Funding Information:
Manuscript received January 27, 2010; revised April 26, 2010; accepted May 4, 2010. Date of publication June 21, 2010; date of current version January 7, 2011. This work was supported by the United States National Science Foundation under Grants 0644511 and 0918856. The Associate Editor coordinating the review process for this paper was Dr. Domenico Grimaldi.

Financiación

Manuscript received January 27, 2010; revised April 26, 2010; accepted May 4, 2010. Date of publication June 21, 2010; date of current version January 7, 2011. This work was supported by the United States National Science Foundation under Grants 0644511 and 0918856. The Associate Editor coordinating the review process for this paper was Dr. Domenico Grimaldi.

FinanciadoresNúmero del financiador
National Science Foundation (NSF)0918856, 0644511

    ASJC Scopus subject areas

    • Instrumentation
    • Electrical and Electronic Engineering

    Huella

    Profundice en los temas de investigación de 'Calibration and field deployment of low-cost fluid flow-rate sensors using a wireless network'. En conjunto forman una huella única.

    Citar esto