Resumen
Background: Delayed lead perforation (occurring more than 1 month after implantation) is a rare complication. Its pathophysiology and optimal management are currently unclear. Methods: Three cases of delayed lead perforation (6-10 month) were identified in patients with low-profile active fixation leads. Results: All cases presented in a subacute fashion with pleuritic chest pain with confirmatory chest x-ray and device interrogation. Given the potential complications of a perforated lead, all cases had the lead extracted under TEE observation with cardiac surgery backup in the operating room. All patients tolerated extraction without complication. Conclusion: Based on these cases, we recommend a management scheme for patients who present with delayed lead perforation.
| Idioma original | English |
|---|---|
| Páginas (desde-hasta) | 251-253 |
| Número de páginas | 3 |
| Publicación | PACE - Pacing and Clinical Electrophysiology |
| Volumen | 28 |
| N.º | 3 |
| DOI | |
| Estado | Published - mar 2005 |
ASJC Scopus subject areas
- Cardiology and Cardiovascular Medicine
Huella
Profundice en los temas de investigación de 'Delayed lead perforation: A disturbing trend'. En conjunto forman una huella única.Citar esto
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver