Dose distribution verification for GYN brachytherapy using EBT Gafchromic film and TG-43 calculation

  • Somayeh Gholami
  • , Hamid Reza Mirzaei
  • , Ali Jabbary Arfaee
  • , Ramin Jaberi
  • , Hassan Ali Nedaie
  • , Seied Rabi Mahdavi
  • , Eftekhar Rajab Bolookat
  • , Ali S. Meigooni

Producción científica: Article › revisión exhaustiva

17 Citas (Scopus)

Resumen

Aim Verification of dose distributions for gynecological (GYN) brachytherapy implants using EBT Gafchromic film. Background One major challenge in brachytherapy is to verify the accuracy of dose distributions calculated by a treatment planning system. Materials and methods A new phantom was designed and fabricated using 90 slabs of 18 cm × 16 cm × 0.2 cm Perspex to accommodate a tandem and Ovoid assembly, which is normally used for GYN brachytherapy treatment. This phantom design allows the use of EBT Gafchromic films for dosimetric verification of GYN implants with a cobalt-60 HDR system or a LDR Cs-137 system. Gafchromic films were exposed using a plan that was designed to deliver 1.5 Gy of dose to 0.5 cm distance from the lateral surface of ovoids from a pair of ovoid assembly that was used for treatment vaginal cuff. For a quantitative analysis of the results for both LDR and HDR systems, the measured dose values at several points of interests were compared with the calculated data from a commercially available treatment planning system. This planning system was utilizing the TG-43 formalism and parameters for calculation of dose distributions around a brachytherapy implant. Results The results of these investigations indicated that the differences between the calculated and measured data at different points were ranging from 2.4% to 3.8% for the LDR Cs-137 and HDR Co-60 systems, respectively. Conclusion The EBT Gafchromic films combined with the newly designed phantom could be utilized for verification of the dose distributions around different GYN implants treated with either LDR or HDR brachytherapy procedures.

Idioma originalEnglish
Páginas (desde-hasta)480-486
Número de páginas7
PublicaciónReports of Practical Oncology and Radiotherapy
Volumen21
N.º5
DOI
EstadoPublished - sept 1 2016

Nota bibliográfica

Publisher Copyright:
© 2016 Greater Poland Cancer Centre

Financiación

I hereby acknowledge Clinical Research Development Unit (CRDU) Shohada Tajrish hospital which was supported this research.

Financiadores
Clinical Research Development Unit

    ASJC Scopus subject areas

    • Oncology
    • Radiology Nuclear Medicine and imaging

    Huella

    Profundice en los temas de investigación de 'Dose distribution verification for GYN brachytherapy using EBT Gafchromic film and TG-43 calculation'. En conjunto forman una huella única.

    Citar esto