Ir directamente a la navegación principal Ir directamente a la búsqueda Ir directamente al contenido principal

Dosimetric characteristics of a new high-intensity 192Ir source for remote afterloading

  • A. S. Meigooni
  • , M. T. Kleiman
  • , J. L. Johnson
  • , D. Mazloomdoost
  • , G. S. Ibbott

Producción científica: Articlerevisión exhaustiva

32 Citas (Scopus)

Resumen

A new high-intensity 192Ir source has recently become commercially available for remote afterloading brachytherapy treatment. The dosimetric characteristics (dose rate constant, radial dose function, and anisotropy function) of this source were experimentally determined through the application of AAPM Task Group 43 recommendations. Complete dosimetric data are presented in this manuscript.

Idioma originalEnglish
Páginas (desde-hasta)2008-2013
Número de páginas6
PublicaciónMedical Physics
Volumen24
N.º12
DOI
EstadoPublished - dic 1997