We have performed a two-part evaluation of assays for β-hydroxy- butyrate (βHB). In Part I, we evaluated the analytical acceptability of a dry film, self-contained analyzer for βHB measurement (KetoSite; GDS Diagnostics, Elkhart, Ind). The assay response is linear and between-day imprecision is acceptable. Acetoacetate (AcAc) interferes severely with this assay. This interference could be minimized by dilution of serum before the assay. With this predilution protocol, KetoSite βHB results compare favorably with an automated Sigma (St. Louis, Mo) βHB assay. In Part II, we compared the clinical utility of βHB measurements (Sigma) with the qualitative measurement of ketones using the nitroprusside test (Acetest; Bayer, Elkhart, Ind) for the diagnosis and management of diabetic ketoacidosis (DKA). Test results for both assays were abnormal for all patients with DKA on admission. However, frequent monitoring with either test during the management of uncomplicated DKA added little if any additional clinical information over routine measurements of serum glucose and total carbon dioxide.
| Idioma original | English |
|---|---|
| Páginas (desde-hasta) | 353-358 |
| Número de páginas | 6 |
| Publicación | American Journal of Clinical Pathology |
| Volumen | 107 |
| N.º | 3 |
| DOI | |
| Estado | Published - mar 1997 |
ODS de las Naciones Unidas
Este resultado contribuye a los siguientes Objetivos de Desarrollo Sostenible
-
Good health and well being
ASJC Scopus subject areas
- Pathology and Forensic Medicine
Huella
Profundice en los temas de investigación de 'Laboratory and clinical evaluation of assays for β-hydroxybutyrate'. En conjunto forman una huella única.Citar esto
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver