Resumen
We describe multiple failed external defibrillation attempts via rescue defibrillation pads for ventricular fibrillation that occurred secondary to direct electrical current transmission through the pericardial sac from electrocautery during robot-assisted internal mammary harvest. Only after resumption of two-lung ventilation and decompression of the iatrogenic pneumothorax was the patient successfully defibrillated. Conditions necessary for robotic intrathoracic surgery may make defibrillation and resuscitation difficult if they become necessary.
| Idioma original | English |
|---|---|
| Páginas (desde-hasta) | 1113-1114 |
| Número de páginas | 2 |
| Publicación | Anesthesia and Analgesia |
| Volumen | 103 |
| N.º | 5 |
| DOI | |
| Estado | Published - nov 2006 |
Financiación
Supported by Department of Anesthesiology, University of Kentucky Chandler Medical Center.