title = "Phantomless patient-specific TomoTherapy QA via delivery performance monitoring and a secondary Monte Carlo dose calculation",
abstract = "Purpose: To describe the validation and implementation of a novel quality assurance (QA) system for TomoTherapy using a Monte Carlo (MC)-based secondary dose calculation and CT detector-based multileaf collimator (MLC) leaf opening time measurement QA verification. This system is capable of detecting plan transfer and delivery errors and evaluating the dosimetric impact of those errors.",
keywords = "Monte Carlo, TomoTherapy, log file, phantomless, quality assurance",