Phantomless patient-specific TomoTherapy QA via delivery performance monitoring and a secondary Monte Carlo dose calculation

  • Lydia L. Handsfield
  • , Ryan Jones
  • , David D. Wilson
  • , Jeffery V. Siebers
  • , Paul W. Read
  • , Quan Chen

Producción científica: Article › revisión exhaustiva

42 Citas (Scopus)

Resumen

Purpose: To describe the validation and implementation of a novel quality assurance (QA) system for TomoTherapy using a Monte Carlo (MC)-based secondary dose calculation and CT detector-based multileaf collimator (MLC) leaf opening time measurement QA verification. This system is capable of detecting plan transfer and delivery errors and evaluating the dosimetric impact of those errors.