Reliability and validity of an accelerometry based measure of static and dynamic postural stability in healthy and active individuals

Producción científica: Article › revisión exhaustiva

62 Citas (Scopus)

Resumen

Postural stability is an important measure in both research and clinical practice. A portable, easy to use device that can provide higher resolution than current clinical tests may allow for better identification of patients or athletes with postural stability deficits. The purpose of this study was to evaluate the ability of a tri-axial accelerometer to quantify postural stability in a healthy athletic population. Ten subjects were recruited to determine the reliability of the accelerometer to measure dynamic postural stability and thirteen were recruited to compare the accelerometer measures across tasks of varying difficulty. Subjects were asked to complete four static postural stability tasks with eyes open and eyes closed and two dynamic postural stability tasks for a total of ten tasks. During each task postural stability was measured using a tri-axial accelerometer and force platform. Differences between postural stability scores between tasks and the correlation between the two measures were assessed. The accelerometer demonstrated moderate to good test-retest reliability (ICC. =. 0.732 to 0.899). Only the medial-lateral axis of the accelerometer showed significant differences between static tasks but all directions were able to show significant differences between static and dynamic tasks. Additionally, Spearman's ranked correlations showed little to no correlation between the accelerometer and force platform scores. Accelerometers are a reliability tool for postural stability that measure low difficulty tasks best in the medial-lateral direction. Low correlation between the accelerometer and force platform suggest that these two methods are not measuring the same components of postural stability.

Idioma originalEnglish
Páginas (desde-hasta)535-539
Número de páginas5
PublicaciónGait and Posture
Volumen41
N.º2
DOI
EstadoPublished - 2015

Nota bibliográfica

Publisher Copyright:
© 2014 Elsevier B.V.

Financiación

The purpose of this study was to evaluate the reliability and validity of an accelerometer-based measure of static and dynamic postural stability. It was hypothesized that an accelerometer-based measure of dynamic postural stability would demonstrate good to excellent test–retest reliability. This hypothesis was supported as the ICC values for the test–retest reliability ranged from 0.732 to 0.899. We also hypothesized that the accelerometer measure of postural stability would detect differences in postural stability between tasks of varying difficulty and display good correlations with the force platform measure of postural stability. This hypothesis was partially supported by the results of this study. The accelerometer measure of postural stability was able to detect differences between tasks of different difficulty; however, these measures did not correlate well with the force platform measure of postural stability.

ASJC Scopus subject areas

  • Biophysics
  • Orthopedics and Sports Medicine
  • Rehabilitation

Huella

Profundice en los temas de investigación de 'Reliability and validity of an accelerometry based measure of static and dynamic postural stability in healthy and active individuals'. En conjunto forman una huella única.

Citar esto