The significance of low titer serum cryptococcal antigen testing from 2017 to 2023 performed in a tertiary care center

  • Thein Myint
  • , Jenna R. Wilson
  • , Mahesh Bhatt
  • , Mark E. Irwin
  • , Julie A. Ribes

Producción científica: Articlerevisión exhaustiva

Resumen

Several false positive lo w serum cryptococcal antigen (SCrAg) reports by lateral flow assay (LFA) were identified in late 2016 at our tertiary care hospital. After the recall and correction of the problem in the reagent, we studied the significance of SCrAg LFA ≤1:10 from January 2017 to October 2023. Of 20 patients with 31 samples of SCrAg LFA ≤1:10, 14 patients (70%) were classified as true positives, four (20%) were indeterminate, and only two (10%) patients w ere false positives. If a new SCrA g LFA ≤1:10 is detected, it should be repeated, and additional workup should be pursued.