Use of error grid analysis to evaluate acceptability of a point of care prothrombin time meter

  • John R. Petersen
  • , Hans M. Vonmarensdorf
  • , Heidi L. Weiss
  • , M. Tarek Elghetany

Producción científica: Articlerevisión exhaustiva

15 Citas (Scopus)

Resumen

Background: Statistical methods (linear regression, correlation analysis, etc.) are frequently employed in comparing methods in the central laboratory (CL). Assessing acceptability of point of care testing (POCT) equipment, however, is more difficult because statistically significant biases may not have an impact on clinical care. We showed how error grid (EG) analysis can be used to evaluate POCT PT INR with the CL. Materials and methods: We compared results from 103 patients seen in an anti-coagulation clinic that were on Coumadin maintenance therapy using fingerstick samples for POCT (Roche CoaguChek XS and S) and citrated venous blood samples for CL (Stago STAR). To compare clinical acceptability of results we developed an EG with zones A, B, C and D. Results: Using 2nd order polynomial equation analysis, POCT results highly correlate with the CL for CoaguChek XS (R2 = 0. 955) and CoaguChek S (R2 = 0. 93), respectively but does not indicate if POCT results are clinically interchangeable with the CL. Using EG it is readily apparent which levels can be considered clinically identical to the CL despite analytical bias. Conclusion: We have demonstrated the usefulness of EG in determining acceptability of POCT PT INR testing and how it can be used to determine cut-offs where differences in POCT results may impact clinical care.

Idioma originalEnglish
Páginas (desde-hasta)131-134
Número de páginas4
PublicaciónClinica Chimica Acta
Volumen411
N.º3-4
DOI
EstadoPublished - feb 2 2010

ASJC Scopus subject areas

  • Biochemistry
  • Clinical Biochemistry
  • Biochemistry, medical

Huella

Profundice en los temas de investigación de 'Use of error grid analysis to evaluate acceptability of a point of care prothrombin time meter'. En conjunto forman una huella única.

Citar esto